Αρχειοθήκη ιστολογίου

Κυριακή 10 Απριλίου 2016

Monte Carlo simulation of semiconductor detector response to 222Rn and 220Rn environments

alertIcon.gif

Publication date: July 2016
Source:Journal of Environmental Radioactivity, Volumes 158–159
Author(s): J. Irlinger, S. Trinkl, M. Wielunksi, J. Tschiersch, W. Rühm
A new electronic radon/thoron monitor employing semiconductor detectors based on a passive diffusion chamber design has been recently developed at the Helmholtz Zentrum München (HMGU). This device allows for acquisition of alpha particle energy spectra, in order to distinguish alpha particles originating from radon and radon progeny decays, as well as those originating from thoron and its progeny decays. A Monte-Carlo application is described which uses the Geant4 toolkit to simulate these alpha particle spectra. Reasonable agreement between measured and simulated spectra were obtained for both 220Rn and 222Rn, in the energy range between 1 and 10 MeV. Measured calibration factors could be reproduced by the simulation, given the uncertainties involved in the measurement and simulation. The simulated alpha particle spectra can now be used to interpret spectra measured in mixed radon/thoron atmospheres. The results agreed well with measurements performed in both radon and thoron gas environments. It is concluded that the developed simulation allows for an accurate prediction of calibration factors and alpha particle energy spectra.



from #MedicinebyAlexandrosSfakianakis via xlomafota13 on Inoreader http://ift.tt/1UT4TJB
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου