The Accuracy of 6 Inexpensive Pulse Oximeters Not Cleared by the Food and Drug Administration: The Possible Global Public Health Implications

imageBACKGROUND: Universal access to pulse oximetry worldwide is often limited by cost and has substantial public health consequences. Low-cost pulse oximeters have become increasingly available with limited regulatory agency oversight. The accuracy of these devices often has not been validated, raising questions about performance. METHODS: The accuracy of 6 low-cost finger pulse oximeters during stable arterial oxygen saturations (SaO2) between 70% and 100% was evaluated in 22 healthy subjects. Oximeters tested were the Contec CMS50DL, Beijing Choice C20, Beijing Choice MD300C23, Starhealth SH-A3, Jumper FPD-500A, and Atlantean SB100 II. Inspired oxygen, nitrogen, and carbon dioxide partial pressures were monitored and adjusted via a partial rebreathing circuit to achieve 10 to 12 stable target SaO2 plateaus between 70% and 100% and PaCO2 values of 35 to 45 mm Hg. Comparisons of pulse oximeter readings (SpO2) with arterial SaO2 (by Radiometer ABL90 and OSM3) were used to calculate bias (SpO2 – SaO2) mean, precision (SD of the bias), and root mean square error (ARMS). RESULTS: Pulse oximeter readings corresponding to 536 blood samples were analyzed. Four of the 6 oximeters tested showed large errors (up to −6.30% mean bias, precision 4.30%, 7.53 ARMS) in estimating saturation when SaO2 was reduced