Αρχειοθήκη ιστολογίου

Κυριακή 8 Οκτωβρίου 2017

Outcomes and reliability of the flow coupler in postoperative monitoring of head and neck free flaps

Objectives/Hypothesis

To assess the accuracy and reliability of the flow coupler relative to the implantable arterial Doppler probe in postoperative monitoring of head and neck free flaps.

Study Design

Retrospective single-institution study, April 2015 to March 2017.

Methods

Both the venous flow coupler and arterial Doppler were employed in 120 consecutive head and neck free flap cases. When Doppler signal loss occurred, flaps were evaluated by physical exam to determine whether signal loss was a true positive necessitating operating room takeback. Sensitivity, specificity, and false positive rate (FPR) were recorded for each device. Logistic regression was conducted to identify user trends over time.

Results

Eleven of 120 patients (9.2%) required takeback, 10 from venous thrombosis and one from arterial thrombosis. Permanent signal loss (PSL) occurred in the flow coupler in all takebacks; PSL occurred in the arterial Doppler only in the case of arterial thrombosis. Salvage rate was 9/11 (81.8%). For the flow coupler, sensitivity was 100%, specificity 86.4%, and FPR 13.6%. For the arterial probe, sensitivity was 9.1%, specificity 97.1%, and FPR 2.9%. A 4.1% decrease in false positives with each additional flow coupler use was observed.

Conclusions

Monitoring the vein via flow coupler has high sensitivity in identifying vascular compromise compared to the arterial probe, especially for venous thrombosis. There is moderate FPR; this decreases with increased usage and, when supplemented with physical examination, does not result in unnecessary takebacks. The flow coupler can be a valuable tool in postoperative monitoring of head and neck free flaps.

Level of Evidence

4. Laryngoscope, 2017



http://ift.tt/2kuc5AU

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου