Αρχειοθήκη ιστολογίου

Παρασκευή 28 Δεκεμβρίου 2018

Bedside ultrasonography for the confirmation of gastric tube placement in the neonate

Yunus Oktay Atalay, Ahmet Veysel Polat, Elif Ozyazici Ozkan, Leman Tomak, Canan Aygun, Joseph Drew Tobias
DOI:10.4103/sja.SJA_413_18  
Background: Naso/Orogastric tube (NOGT) misplacement can lead to significant complications. Therefore, the assessment of tube position is essential to ensure patient safety. Although radiography is considered the gold standard for determining NOGT location, new methods may be helpful in reducing repetitive radiation exposure, especially for neonates. In this study, we sought to investigate if bedside ultrasonography (BUSG) can be used to verify NOGT placement in neonatal intensive care patients. Materials and Methods: Infants requiring NOGT placement were enrolled. After insertion of the NOGT, the location was first identified using BUSG and then confirmed using abdominal radiography for comparison. Results: The study cohort included 51 infants with an average gestational age of 34 ± 4.9 weeks. BUSG determined the NOGT location correctly with a sensitivity of 92.2%. The location of the NOGT could not be determined by BUSG in four neonates (7.8%). In one infant, the NOGT was positioned in the esophagus, as determined both by BUSG and radiography. Conclusion: BUSG is a promising diagnostic tool for determining NOGT location in neonates, thereby eliminating the need for abdominal radiography.

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου