Αρχειοθήκη ιστολογίου

Πέμπτη 30 Ιουνίου 2016

SU-F-T-152: Experimental Validation and Calculation Benchmark for a Commercial Monte Carlo Pencil BeamScanning Proton Therapy Treatment Planning System in Heterogeneous Media

Purpose:

Eclipse AcurosPT 13.7, the first commercial Monte Carlo pencil beam scanning (PBS) proton therapy treatment planning system (TPS), was experimentally validated for an IBA dedicated PBS nozzle in the CIRS 002LFC thoracic phantom.

Methods:

A two-stage procedure involving the use of TOPAS 1.3 simulations was performed. First, Geant4-based TOPAS simulations in this phantom were experimentally validated for single and multi-spot profiles at several depths for 100, 115, 150, 180, 210 and 225 MeV proton beams, using the combination of a Lynx scintillation detector and a MatriXXPT ionization chamber array. Second, benchmark calculations were performed with both AcurosPT and TOPAS in a phantom identical to the CIRS 002LFC, with the exception that the CIRS bone/mediastinum/lung tissues were replaced with similar tissues that are predefined in AcurosPT (a limitation of this system which necessitates the two stage procedure).

Results:

Spot sigmas measured in tissue were in agreement within 0.2 mm of TOPAS simulation for all six energies, while AcurosPT was consistently found to have larger spot sigma (calibration by MatriXXPT, the agreements between profiles measurements and TOPAS simulation, and calculation benchmarks are over 97% except near the end of range using 2 mm/2% gamma criteria. Overdosing and underdosing were observed at the low and high density side of tissue interfaces, respectively, and these increased with increasing depth and decreasing energy. Near the mediastinum/lung interface, the magnitude can exceed 5 mm/10%. Furthermore, we observed >5% quenching effect in the conversion of Lynx measurements to dose.

Conclusion:

We recommend the use of an ionization chamber array in combination with the scintillation detector to measure absolute dose and relative PBS spot characteristics. We also recommend the use of an independent Monte Carlo calculation benchmark for the commissioning of a commercial TPS.

Partially supported by Varian Medical System under the master agreement between Varian and University of pennsylvania



from #MedicinebyAlexandrosSfakianakis via xlomafota13 on Inoreader http://ift.tt/294ZWIL
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου