Αρχειοθήκη ιστολογίου

Δευτέρα 12 Νοεμβρίου 2018

Intravascular Cooling Device Versus Esophageal Heat Exchanger for Mild Therapeutic Hypothermia in an Experimental Setting

BACKGROUND: Targeted temperature management is a standard therapy for unconscious survivors of cardiac arrest. To date, multiple cooling methods are available including invasive intravascular cooling devices (IVDs), which are widely used in the clinical setting. Recently, esophageal heat exchangers (EHEs) have been developed providing cooling via the esophagus that is located close to the aorta and inferior vena cava. The objective was to compare mean cooling rates, as well as differences, to target temperature during maintenance and the rewarming period of IVD and EHE. METHODS: The study was conducted in 16 female domestic pigs. After randomization to either IVD or EHE (n = 8/group), core body temperature was reduced to 33°C. After 24 hours of maintenance (33°C), animals were rewarmed using a target rate of 0.25°C/h for 10 hours. All cooling phases were steered by a closed-loop feedback system between the internal jugular vein and the chiller. After euthanasia, laryngeal and esophageal tissue was harvested for histopathological examination. RESULTS: Mean cooling rates (4.0°C/h ± 0.4°C/h for IVD and 2.4°C/h ± 0.3°C/h for EHE; P

https://ift.tt/2DcGTOB

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου